From: Histological quantification of the tissue damage caused by PlasmaJet™ coagulator
Power level | 2 s | 5 s | ||
---|---|---|---|---|
Depth (mm) | Width (mm) | Depth (mm) | Width (mm) | |
10% | 0.48 ± 0.16 | 4.45 ± 0.82 | 0.52 ± 0.18 | 5.01 ± 0.69 |
15% | 0.44 ± 0.14 | 4.25 ± 0.62 | 0.50 ± 0.13 | 4.56 ± 0.66 |
20% | 0.46 ± 0.14 | 4.05 ± 0.54 | 0.51 ± 0.11 | 4.82 ± 0.76 |