Skip to main content

Table 3 Mean ± SD depth and width (lateral spread) of tissue damage caused by PlasmaJetTM coagulator in the fallopian tube in vitro

From: Histological quantification of the tissue damage caused by PlasmaJet™ coagulator

Power level

2 s

5 s

Depth (mm)

Width (mm)

Depth (mm)

Width (mm)

10%

0.49 ± 0.13

5.45 ± 0.78

0.52 ± 0.15

5.25 ± 0.79

15%

0.50 ± 0.15

4.65 ± 0.72

0.45 ± 0.12

4.96 ± 0.70

20%

0.49 ± 0.18

4.90 ± 0.55

0.48 ± 0.09

4.80 ± 0.56