Skip to main content

Advertisement

Table 4 Mean ± SD depth and width of tissue damage caused by PlasmaJetTM coagulator in vivo

From: Histological quantification of the tissue damage caused by PlasmaJet™ coagulator

  Depth (mm) mean ± SD Width (mm) mean ± SD
Uterus (n = 15) 0.63 ± 0.19 4.66 ± 0.05
Ovary (n = 10) 0.61 ± 0.14 4.05 ± 0.61
Fallopian tube (n = 10) 0.63 ± 0.18 4.51 ± 0.77