The development and psychometric validation of a low-cost anthropomorphic 3D-printout simulator for training basic skills applicable to office-based hysteroscopy

Abstract

Background

Hysteroscopy training requires the development of specific psychomotor skills. Few validated low-cost models exist in hysteroscopy. The main objective of this study is to determine face, content, and construct validity of a simulator designed for training basic hysteroscopy skills applied to office-based hysteroscopy.

Methods

Twenty-five hysteroscopy experts and 30 gynecology residents participated in this prospective observational study. The simulator consisted of three color-textured, silicone-coated anthropomorphic 3-dimensional (3-D) printout uterine models inside a box. Each uterine model in the simulator was designed to develop one of the following basic hysteroscopic skills: hysteroscopic navigation, direct biopsy, and foreign body removal. Participants performed five video-recorded simulation attempts on each model. Procedure-specific checklists were used to rate performance.

Results

Median scores (25th–75th percentiles; p-value) 4 (3–4; p < 0.001) and surgical experience associated to the simulated procedures 4 (3–4; p < 0.001) indicated positive perceptions as to the realism of the internal cavity of the simulator. Median scores of 4 (3–4; p < 0.001) were assigned to the realism and utility of the tasks performed in the simulator for enhancing novice training in hysteroscopy. Expert performance scores were significantly higher and task completion times were significantly lower than those of novices in the navigation exercise (F(1,53) = 56.66; p < 0.001), the directed biopsy exercise (F(1,53) = 22.45; p < 0.001), and the foreign body removal exercise (F(1,53) = 58.51; p < 0.001). Novices’ performance improved on all three exercises: navigation exercise (F(1,53) = 182.44; p < 0.001), directed endometrial biopsy (F(1,53) = 110.53; p < 0.001), and foreign body removal (F(1,53) = 58.62; p < 0.001). Experts’ task completion times were significantly lower when compared to that of novices in the five attempts (p < 0,001) of the exercises: navigation (F(1,48) = 25.46; p < 0.001), directed biopsy (F(1,46) = 31.20; p < 0.001), and foreign body removal (F(1,50) = 69.8; p < 0.001). Novices’ task completion times diminished significantly throughout the sequence of exercises.

Conclusions

The low-cost simulator designed for the acquisition of basic skills in hysteroscopy demonstrated face, content, and construct validity.

Background

Hysteroscopy is a minimally invasive procedure considered the gold standard for the evaluation and treatment of intrauterine diseases [1]. Currently available, small-diameter scopes have permitted the performance of in-office hysteroscopic procedures, allowing non-traumatic insertion into the cervix, a vaginoscopy technique [2, 3]. Therefore, hysteroscopic proficiency requires specific training for the development of new psychomotor skills. The importance of a simulator as a training tool has been extensively emphasized [4, 5].

Simulators allow training in a safe environment, real-time expert feedback and repeated practice of each step of the procedure [6,7,8]. Recommendations have been made for institutions to provide dry labs for training endoscopic surgical skills [9, 10].

The successful integration of a new simulation system into a training curriculum demands rigorous evaluation of its validity [11]. The first step is to demonstrate that new simulator resembles real-life situations, that is, face validity. The extent to which new simulator allows acquisition of abilities required for proficiently performing in real patients, content validity, must also be assured. Thereon, validity procedures aim to demonstrate that new simulator allows detection of distinct levels of operator proficiency, for example, performance of novices and experts [12]. It is also desirable to demonstrate that simulator allows detection of changes in performance, learning curve, during repeated attempts [13].

We hypothesized that a novel, inorganic, anthropomorphic simulator incorporates 3-D printing would demonstrate high levels of face, content, and construct validity. The present study aimed to test these hypotheses by applying specific procedures for face, content, and construct validation.

Materials and methods

Data were collected at two general hospitals. The IRB (institutional review board) approved the study at the participant institutions. Written informed consent was obtained from participants before inclusion in the study. From October 2018 to April 2019, a convenience sample consisting of 30 gynecology residents with no previous experience in hysteroscopy and 25 experts participated in the study. Experts were gynecologic surgeons who had received formal training in hysteroscopic procedures and had been practicing diagnostic and operative hysteroscopy for more than 3 years at the time they were enrolled in the study.

The simulator

The anthropomorphic anatomical models used in the simulator were manufactured in acrylonitrile butadiene styrene (ABS) on a fusion deposition modeling (FDM) 3-D printer (3-D Printer Prusa I3 Rework, ITCTERM, Brazil). The 3-D model was scanned from a 9 × 7 × 6 cm pyriform mold simulating a human uterus. The models contained two hemicavities internally lined with textured, rose-colored acetic silicone to mimic the appearance of the human endometrium. A 2-cm diameter opening was created to allow the atraumatic passage of the hysteroscope.

Three uterine models were embedded in latex foam inside an opaque 30-cm-wide, 28-cm-long, 18-cm-high cardboard box. Three 3-cm circular apertures were made on the longest box wall. The three apertures were framed with ABS ring-shaped connectors to which red-colored male condoms were attached and connected to the opening of the uterine models inside the box simulating vaginas (Fig. 1). Each uterine model in the simulator was designed to develop one of the basic hysteroscopic skills.

Fig. 1
figure 1

The simulator. An internal view of the simulator with the three uterine models embedded in a foam base and fixed with Velcro straps is shown on the left panel. Male condoms connect the aperture of the model to the wall of the simulator box. From left to right, the uterine models designed to the exercises of navigation, direct biopsy, and foreign body removal are shown. Sensor wires in the second model and balloons inserted through the wall of the third model can also be seen. Images of the internal surface of the models intended for training navigation, direct biopsy, and foreign body extraction are shown from top to bottom in the right panel of the figure

The first model was designed to develop intrauterine navigation skills. It was equipped with nine distinct adhesive images positioned on the anterior, posterior, left lateral, and right lateral walls; on the uterine fundus; and on the anterior and posterior tubal ostia, represented by two red markings inside the model.

The second uterine model was designed to develop direct biopsy skills. Five electronic touch-sensors (Micro Switch, KW-1 3pins, Dongnan, China) covered by a non-slip adhesive tape were connected to a microcontroller board, Arduino Uno (Smarts Projects, Ivrea, Italy) which was connected to a microcomputer via a USB (universal serial bus) port. The sensors were positioned through openings in the model’s wall in the uterine fundus and on the anterior, posterior, left, and right walls of the inner surface of the model. A visual sign on the computer screen indicated which sensor had been pressed at each attempt.

The third uterine model was designed for developing foreign body removal skills. The model was equipped with five 8-mm-diameter apertures on the model wall through which 8-mm-diameter, 4-cm-long mini-balloons were inserted 5 mm into the model cavity.

Face and content validity

Face and content validities were based on expert perceptions of the realism of tasks performed on the simulator compared to procedures performed on real patients. Face validity was evaluated by assessing the degree of expert concordance with the following statements: (1) the appearance of the simulator’s internal cavity resembles that of the human uterine body cavity, and (2) the procedures performed on the simulator resemble the office-based hysteroscopy procedures performed on human patients. Responses were measured on four-point unidirectional, forced-response scales, where the extreme points 1 and 4 indicated full disagreement and full agreement, respectively. Also, the realism of the global experience in performing the tasks on the simulator was assessed on a 10-point rating scale (1 = absolutely non-realistic; 10 = absolutely realistic).

Content validity was based on experts’ agreement with statements addressing the similarity between the tasks performed on the simulator, the hysteroscopic maneuvers performed in real patients, for example hand-eye coordination, accurate visualization through a 30° angle scope and the use of grasping forceps. Also, the utility of learning experiences in the simulator for training novices was used to assess the content validity of the simulator. Responses were measured on four-point unidirectional scales where the extreme points 1 and 4 indicated full disagreement and full agreement, respectively.

Construct validity

The ability to discriminate between expert and resident performance determined the construct validity of the simulator. Participants’ performance was measured on procedure-specific checklists used for rating performance on the video-recorded simulation sessions.

For developing the procedure-specific checklists (Table 2 in Appendix), three authors (A.R.P., L.B.G., and L.K.V.) independently searched the literature to create technical performance checklists addressing the main steps of the simulation tasks: hysteroscopic navigation, direct biopsy, and foreign body removal [14, 15]. Items in the checklists addressed ergonomics, image visualization, safe navigation, and handling the grasping forceps [12]. Items of the checklists represented the elements or steps identified in the three tasks that could be measured and potentially differentiate among levels of technical competence. Three-point scales were added to the scoring rubrics, where zero indicated an unskilled performance, as expected from a novice; one indicated a somewhat skilled performance, as expected from an in-training novice; and two indicated a skilled performance, as expected from an experienced surgeon [16].

Simulation sessions

Simulation sessions took place in private rooms equipped with a video endoscopy equipment consisting of a high-definition monitor (Karl Storz, Germany), a cold light source XENON 300 (Karl Storz, Germany), and a video-camera IMAGE 1 HUB™ (Karl Storz, Germany). Also available in the simulation room were HOPKINS® Forward-Oblique 30°, 2.9-mm Telescope (Karl Storz, Germany), BETTOCCHI® inner and outer sheaths (Karl Storz, Germany), and semirigid 5-Fr grasping forceps (Karl Storz, Germany), which were used to complete the tasks in the simulator. The simulator box was positioned on a 90-cm-high table for the training sessions. The study participants were offered the choice of performing the tasks in the standing or the sitting position. Most trainees preferred the sitting position with the video equipment positioned anteriorly and to their left, allowing a full view of the screen. The same investigator (A.R.P.) conducted all simulation sessions, which involved only one participant at a time. Formative feedback on participant performance was allowed [13].

Simulation sessions consisted of a preparatory and a hands-on phase. The preparatory phase aimed to provide the participant with relevant theoretical and practical guidance about the tasks to be simulated. The preparatory phase started soon after the arrival of the participant at the simulation room. The instructor started by providing a structured 20-min presentation addressing the definition, the indications, the equipment, the technique, and the complications of office-based hysteroscopy. Next, participants were informed about the technical aspects of the tasks to be simulated, their metrics, and goals. Participants were further instructed about how to handle the scope and the grasping forceps. The preparatory phase of the experiment finished after the instructor presented the simulator, demonstrated the tasks, and allowed the participants to manipulate the simulator, the video, and the hysteroscopy equipment for approximately 5 min before starting the hands-on phase of the simulation session.

Participants performed three exercises (tasks) during the hands-on phase of the simulation session. The navigation exercise (task 1) consisted of identifying, visualizing, and centering the nine targets and simulated tubal ostia inside the first model followed by obtaining a panoramic view of the interior uterine model, before withdrawing the hysteroscope. The aim of the direct biopsy simulation exercise (task 2) was to press the sensors positioned inside model 2 with the tip of a grasping forceps. The foreign body removal simulation exercises (task 3) consisted of using grasping forceps to clamp and pull the mini-balloons inserted on the wall of the third model at least 3 cm into the simulator cavity. Participants repeated each exercise five times. Task completion times were counted from the moment of insertion to the moment of withdrawal of the hysteroscope from the simulator. Participant performance was digitally recorded (Canon EOS 60D 18 MP, Canon U.S.A., Inc., Huntington, NY, USA) for further analyses. Three authors (A.R.P., L.B.G., and L.K.V.) rated the videos independently [17].

At the end of the simulation session, participants answered the demographic data questionnaire. Experts also completed the eight-item face and content validity questionnaire.

Statistical analysis

Statistical analysis was performed utilizing SPSS v 26.0 (SPSS Inc., Chicago, IL, USA). Basic descriptive statistics were calculated for demographic data. All continuous variables were presented as mean (standard deviation) or median (range), and nominal variables were presented as frequency (percentage). Kolmogorov-Smirnov test was used to assess normality. Generalized linear mixed-model (GLM mixed-model) ANOVA was used to compare scores and task completion times within and between groups. Median scores of the items used to assess the face and content validities were tested against the mathematical center of the scales by using one-sample Wilcoxon signed-rank tests.

Psychometric analyses of the technical performance checklists included exploratory factor analyses by the principal-component extraction method and varimax rotation with Kaiser normalization to determine the factorial structure of the instruments and the estimation of Cronbach’s alpha coefficients to assess the internal consistency of the technical performance checklists.

Inter-rater agreement was evaluated by estimating intraclass correlation coefficients and their 95% confidence intervals among the scores attributed by raters 1, 2, and 3 to the items in the checklists and among the average scores, under the following assumptions: the same raters assessed all videotaped procedures, the selected raters were representatives of a larger sample of raters (experts in ambulatorial hysteroscopy), the reliability of the mean value of multiple raters was the measure of interest, and we searched for consistency of ratings. Based on the assumptions mentioned above, two-way random-effects models were used [18].

The estimated sample size was 44 participants for five attempts with a power of 80% at an α level of 0.05. Calculations were based on the effect size (ε2) equal to 0.057 observed in a previous study [12]. Fifty-five subjects were enrolled in the study to account for eventual losses.

Results

Twenty-five experts and thirty residents completed the study. Demographic data are shown in Table 1. A total of 825 videos records of participants’ performance were obtained and randomized (www.ramdom.org).

Table 1 Demographic data of the study subjects

Face and content validity

Median scores (25th–75th percentiles) assigned by experts to the statements that assessed face validity were significantly higher than the mathematical center of the scale (p < .001), indicating positive perceptions about the realism of the internal cavity of the simulator 4 (3–4; p < .001) and of the surgical experience associated the simulated procedures 4 (3–4; p < .001). Also, the median score of the global experience on the simulator was 9 (25th–75th percentiles = 9–10), differing significantly from the mathematical center of the scale (p < .001). Median scores of 4 (3–4; p < .001) were assigned to the realism and utility of the tasks performed in the simulator as applicable to the training of novices in hysteroscopy.

Psychometric analyses of the technical performance procedure-specific checklists

A total of 825 video recordings of participants’ performance were obtained, from which 27 (3.2%) were excluded due to the low quality of images. Three authors rated 798 videos. Ratings were used for psychometric analyses of the technical performance checklists.

The factorial structure of the checklist used to assess participants’ performance in task 1 comprised two factors, which explained 55.89% of score variance: factor 1 (items related to image visualization ability) (Eigenvalue = 7.42; explained variance = 49.48%) and factor 2 (ergonomics-related items) (Eigenvalue = 1.56; explained variance = 10.41%). The factorial structure of the scale used to assess participants’ performance in task 2 comprised a single factor that explained 51.62% of score variance (Eigenvalue = 3.097). The factorial structure of the scale used to assess participants’ performance in task 3 comprised a single factor that explained 52.92% of score variance (Eigenvalue = 3.176).

Cronbach’s alpha coefficients of the checklist designed to assess task 1 (navigation) were 0.92 (95% confidence interval, CI = 0.91–0.93), 0.77 (95% CI, 0.75–0.80), and 0.90 (95% CI, 0.89–0.91) for image visualization ability factors, ergonomics, and for the global scale, respectively. Based on these findings, the global score was chosen to measure participants’ performance on the navigation tasks. The Cronbach’s alpha coefficient estimated for the checklist designed to assess performance at tasks 2 and 3 were 0.80 (95% CI = 0.78–0.83) and 0.82 (95% CI = 0.80–0.84), respectively.

Intraclass correlation coefficients among raters’ average scores on items addressing image visualization and ergonomics in task 1 checklist were 0.55 (95% CI = 0.45–0.64) and 0.76 (95% CI = 0.71–0.81). Intraclass correlation coefficients among raters’ average scores were 0.75 (95% CI = 0.69–0.79) and 0.75 (95% CI = 0.69–0.80) for tasks 2 and 3 technical performance checklists, respectively.

Construct validity

Expert technical performance scores were constant during the five attempts and significantly higher than novices for navigation tasks (F(1,53) = 56.66; p < .001; Fig. 2a); direct endometrial biopsy (F(1,53) = 22.45; p < .001; Fig. 3a); and foreign body removal (F(1,53) = 58.51; p < .001; Fig. 4a). Residents’ performance at the navigation tasks improved from the third to the fifth attempt (F(1,53) = 182.44; p < .001; Fig. 2a). Significant improvement was observed at the fourth attempt at the direct endometrial biopsy (F(1,53) = 110.53; p < .001; Fig. 3a). Residents’ performance improved significantly from the second through the fifth attempts at the foreign body removal exercise (F(1,53) = 58.62; p < .001; Fig. 4a).

Fig. 2
figure 2

Navigation task. a Global performance scores. b Task completion time. Brackets and the respective vertically oriented p-values refer to between-group comparisons. Horizontally oriented p-values refer to within-group comparisons to the first attempt

Fig. 3
figure 3

Direct biopsy task. a Global performance scores. b Task completion time. Brackets and the respective vertically oriented p-values refer to between-group comparisons. Horizontally oriented p-values refer to within-group comparisons to the first attempt

Fig. 4
figure 4

Foreign body removal task. a Global performance scores. b Task completion time. Brackets and the respective vertically oriented p-values refer to between-group comparisons. Horizontally oriented p-values refer to within-group comparisons to the first attempt

Experts’ times to complete the three tasks were significantly lower than those of residents at all attempts at navigation (F(1,48) = 25.46; p < .001), direct biopsy (F(1,46) = 31.20; p < .001), and foreign body removal task (F(1,50) = 69.8; p < .001) as shown in Fig. 4b. Experts’ and residents’ times to perform the navigation task decreased significantly throughout the five attempts (F(1,48) = 30.98; p < .001; Fig. 2b). Residents’ times decreased significantly from the second through the fifth attempts during the sequence of simulated endometrial biopsy (F(1,46) = 4.35; p < .001; Fig. 3b). At the foreign body removal task, residents’ times decreased significantly from the second through the fifth attempts (F(1,50) = 5.85; p < .019; Fig. 4b), while experts’ times decreased significantly from the third to the fifth attempt (F(1,46) = 4.35; p < .001) and (F(1,50) = 5.85; p < .019; Fig. 4b).

Discussion

The main results of this study were that the new simulator exhibited high face, content, and construct validity. Also, highly reliable instruments were used to produce the measures of technical performance.

Face validity of anthropomorphic simulators requires close resemblance between what the operator sees in the simulator and in real patients (realism). For this reason, the face validity of anthropomorphic simulators must rely on expert perception about the realism of the internal appearance, of the simulator [19, 20]. To meet the requirement, the simulator developed for this present study was lined with colored texturized silicone to give it a realistic aspect, mimicking the endometrial cavity. The silicone lining also aimed to protect the hysteroscope lens during the training sessions. The tubal ostia were marked in red, standing out from the rest of the cavity to direct navigation and differentiating this simulator from other low-fidelity alternatives, such as the simulator developed by the European Academy of Gynaecological Surgery [21] and the hysteroscopic component of the Essentials in Minimally Invasive Gynecology Hysteroscopy Simulation System (EMIG) [22]. The present simulator shares high face validity as hysteroscopy component EMIG system recently developed during this research study [22]. Some fruits and vegetables can also be used as inexpensive simulators that resemble the human uterus and allow training surgical procedures [5], such as the butternut pumpkin model, which was deemed an excellent choice and used by the Royal Australian and New Zealand College of Obstetrics and Gynecology [23]. However, such models are perishable and may not be used in hospital settings.

High content validity indicates that the simulator allows the acquisition of the basic skills necessary to safely and proficiently complete the addressed procedures. Content validity grants the simulator the property of serving as a useful training tool. Office-based hysteroscopic experience is necessary to assess content validity; only experts’ ratings were used in this study, as preconized elsewhere [20, 24]. Because the simulator includes three procedure-specific uterine models, users can acquire skills in the most frequent and challenging maneuvers performed during office-based hysteroscopy. Training navigation includes the manipulation of the hysteroscope within the uterine cavity to obtain a clear view through the 30°-angle optics; novices can also train the reaching of small targets inside the uterine cavity as required for the successful performance of hysteroscopic endometrial biopsy. The simulator also permits the acquisition of skills at manipulating grasping forceps, which are crucial for foreign body removal. Furthermore, different from virtual simulators, our simulator provides haptic feedback that increases the realism of the task [25]. To the authors’ knowledge, no other study addressed the content validity of simulators designed for the acquisition of skills for office-based hysteroscopy [5, 26].

High construct validity of a simulator indicates the ability to differentiate the performance of experts and novices practicing on it. Besides, the demonstration that training in the simulator improves skills over time is also an indicator of construct validity, which is a necessary feature of simulators designed to follow the learning curve of novices during their path to proficiency [27]. By strategically positioning targets on every uterine wall, the simulator posed distinct levels of difficulty to assess the targets, thus allowing for differentiation between expert and novice performance. Furthermore, the success at progressively more difficult targets progressively increased the summative score, allowing for the detection of technical performance improvement among novices. Similarly, progressively decreased time to touch the targets might also be evidence of performance improvement, as demonstrated in a previous study [28]. Although significant improvement in technical ability and decreases in task completion time during five attempts, residents did not reach experts’ performance level, indicating that learning curves of the simulated tasks are longer and deserve further studies. However, this finding corroborates the need for hysteroscopic simulation training before a performance in real patients [5, 29, 30].

The strengths of the current simulator are the high face, content and construct validity, and its affordable production cost (approximately US$150.00), easy reproducibility, and portability. These characteristics make the simulator an attractive tool for training and diffusing office-based hysteroscopy technique in developing countries where scarce resources challenge the acquisition of expensive simulators [31].

Low cost and easy reproducibility were the main concerns in the project of the simulator. Incorporating 3-D technology into the manufacturing of the uterine models fulfilled both requirements. However, equipping the model with resources for training vaginoscopy, the insertion of the hysteroscope through the uterine cervix and managing irrigation and complications demonstrated in virtual simulators would have substantially increased the cost of the simulator. These limitations have also been found in other low-fidelity simulation studies such as HYSTT and hysteroscopic component of EMIG [21, 22, 32]. Still, low-fidelity simulators demonstrate training capacity for basic skills [32].

Conclusion

Before its implementation as a training tool, a newly developed simulator requires independent validation of it face, content, and construct validity. This study described a new simulator designed for the acquisition of skills relevant to the safe performance of office-based hysteroscopic procedures. Training on the simulator produced realistic experiences for the user, an indication of high face validity; experts also highly rated the usefulness of the exercises performed on the simulator for the training of novices, an indication of content validity. Performance on the simulator also differentiates levels of surgical experience, an indication of construct validity.

Availability of data and materials

Data will be made available on request.

Abbreviations

3-D:

Three dimensional

p :

P-value

F :

Factor; factor analysis

IRB:

Institutional review board

ABS:

Acrylonitrile butadiene styrene

FDM:

Fusion deposition modeling

USB:

Universal Serial Bus

GLM mixed-model:

Generalized linear mixed model

ε 2 :

Effect size

CI:

Confidence interval

References

  1. Salazar CA, Isaacson KB (2018) Office operative hysteroscopy: an update. J Minim Invasive Gynecol 25(2):199–208. https://doi.org/10.1016/j.jmig.2017.08.009

    Google Scholar 

  2. Bettocchi S, Nappi L, Ceci O et al (2004) Office hysteroscopy. Gynecol Clin North Am 31:641–654. https://doi.org/10.1016/j.ogc.2004.05.007

    Google Scholar 

  3. Di Spiezio SA, Bettocchi S, Spinelli M et al (2010) Review of new office-based hysteroscopic procedures 2003-2009. J Minim Invasive Gynecol 17(4):436–448. https://doi.org/10.1016/j.jmig.2010.03.014

    Google Scholar 

  4. Nappi L, Greco P, Ciavarella M, Angioni S, Becker S, Wattiez A (2013) Re: Intensive programme in ‘minimally invasive procedures in endoscopic gynecology’. A teaching course supported by EU funds. J Minim Invasive Gynecol 20(1):133–134

    Google Scholar 

  5. Gambadauro P, Milenkovic M, Hadlaczky G (2018) Simulation for training and assessment in hysteroscopy: a systematic review. J Minim Invasive Gynecol 25(6):963–973. https://doi.org/10.1016/j.jmig.2018.03.024

    Google Scholar 

  6. Ericsson KA, Krampe RT, Tesch-Römer C (1993) The role of deliberate practice in the acquisition of expert performance. Psychol Rev 100(3):363–406. https://doi.org/10.1037/0033-295x.100.3.363

    Google Scholar 

  7. Rackow BW, Solnik MJ, Tu FF, Senapati S, Pozolo KE, Du H (2012) Deliberate practice improves obstetrics and gynecology residents’ hysteroscopy skills. J Grad Med Educ 4(3):329–334. https://doi.org/10.4300/JGME-D-11-00077.1

    Google Scholar 

  8. Munro MG (2012) Surgical simulation : where have we come from? Where are we now? Where are we going? J Minim Invasive Gynecol 19(3):272–283. https://doi.org/10.1016/j.jmig.2012.01.012

    Google Scholar 

  9. Campo R, Puga M, Meier FR, Wattiez A, De Wilde RL (2014) Excellence needs training “Certified programme in endoscopic surgery”. Facts, Views Vis ObGyn 6(4):240–244

    Google Scholar 

  10. Bharathan R, Setchell T, Miskry T, Darzi A, Aggarwal R (2014) Gynecologic endoscopy skills training and assessment: review. J Minim Invasive Gynecol 21(1):28–43. https://doi.org/10.1016/j.jmig.2013.07.016

    Google Scholar 

  11. Gallagher AG, Ritter EM, Satava RM (2003) Fundamental principles of validation, and reliability: rigorous science for the assessment of surgical education and training. Surg Endosc Other Interv Tech 17(10):1525–1529. https://doi.org/10.1007/s00464-003-0035-4

    Google Scholar 

  12. Ml B, Tuchschmid S, Fink D, Székely G, Harders M (2010) Establishing construct validity of a virtual-reality training simulator for hysteroscopy via a multimetric scoring system. Surg Endosc Other Interv Tech 24(1):79–88. https://doi.org/10.1007/s00464-009-0582-4

    Google Scholar 

  13. De Oliveira Filho GR, Mettrau FAC (2018) The effect of high-frequency, structured expert feedback on the learning curves of basic interventional ultrasound skills applied to regional anesthesia. Anesth Analg 126(3):1028–1034. https://doi.org/10.1213/ANE.0000000000002748

    Google Scholar 

  14. Tuchschmid S, Bajka M, Bachofen D, Székely G, Harders M (2007) Objective surgical performance assessment for virtual hysteroscopy. Stud Health Technol Inform 125:473–478

    Google Scholar 

  15. Zevin B, Levy JS, Satava RM, Grantcharov TP (2012) A consensus-based framework for design, validation, and implementation of simulation-based training curricula in surgery. J Am Coll Surg 215(4):580–586.e3. https://doi.org/10.1016/j.jamcollsurg.2012.05.035

    Google Scholar 

  16. Jacoby J, Michael SM (2014) Three-point Likert scales. J Mark Res 8(4):495–500

    Google Scholar 

  17. Burchard ER, Lockrow EG, Zahn CM, Dunlow SG, Satin AJ (2007) Simulation training improves resident performance in operative hysteroscopic resection techniques. Am J Obstet Gynecol 197(5):542.e1–542.e4. https://doi.org/10.1016/j.ajog.2007.07.023

    Google Scholar 

  18. Koo TK, Li MY (2016) A guideline of selecting and reporting intraclass correlation coefficients for reliability research. J Chiropr Med 15(2):155–163. https://doi.org/10.1016/j.jcm.2016.02.012

    Google Scholar 

  19. Bright E, Vine S, Wilson MR, Masters RSW, McGrath JS (2012) Face validity, construct validity and training benefits of a virtual reality turp simulator. Int J Surg 10(3):163–166. https://doi.org/10.1016/j.ijsu.2012.02.012

    Google Scholar 

  20. Dulan G, Rege RV, Hogg DC, Gilberg-Fisher KK, Tesfay ST, Scott DJ (2012) Content and face validity of a comprehensive robotic skills training program for general surgery, urology, and gynecology. Am J Surg 203(4):535–539. https://doi.org/10.1016/j.amjsurg.2011.09.021

    Google Scholar 

  21. Janse JA, Tolman CJ, Veersema S, Broekmans FJM, Schreuder HWR (2014) Hysteroscopy training and learning curve of 30° camera navigation on a new box trainer: the HYSTT. Gynecol Surg 11(2):67–73. https://doi.org/10.1007/s10397-014-0833-9

    Google Scholar 

  22. Munro MG, Brown AN, Saadat S et al (2019) Essentials in minimally invasive gynecology manual skills pilot validation trial. J Minim Invasive Gynecol. https://doi.org/10.1016/j.jmig.2019.04.018

  23. Kingston A, Abbott J, Lenart M, Vancaillie T (2004) Hysteroscopic training: the butternut pumpkin model. J Am Assoc Gynecol Laparosc 11(2):256–261. https://doi.org/10.1016/S1074-3804(05)60210-1

    Google Scholar 

  24. De Oliveira Filho GR, Helayel PE, Conceição DB, Garzel IS, Pavei P, Ceccon MS (2008) Learning curves and mathematical models for interventional ultrasound basic skills. Anesth Analg 106(2):568–573. https://doi.org/10.1213/ane.0b013e3181605412

    Google Scholar 

  25. Verdaasdonk EGG, Stassen LPS, Monteny LJ, Dankelman J (2006) Validation of a new basic virtual reality simulator for training of basic endoscopic skills: The SIMENDO. Surg Endosc Other Interv Tech 20(3):511–518

    Google Scholar 

  26. Savran MM, Maya S, Sørensen D et al (2016) Training and assessment of hysteroscopic skills: a systematic review. J Surg Educ 73(5):906–918. https://doi.org/10.1016/j.jsurg.2016.04.006

    Google Scholar 

  27. Mokkink LB, Terwee CB, Gibbons E et al (2010) Inter-rater agreement and reliability of the COSMIN (consensus-based standards for the selection of health status measurement instruments) checklist. BMC Med Res Methodol 10(1):82. https://doi.org/10.1186/1471-2288-10-82

    Google Scholar 

  28. Alici F, Buerkle B, Tempfer CB (2014) Objective Structured Assessment of Technical Skills (OSATS) evaluation of hysteroscopy training: a prospective study. Eur J Obstet Gynecol Reprod Biol 178:1–5. https://doi.org/10.1016/j.ejogrb.2014.04.032

    Google Scholar 

  29. Janse JA, Driessen SRCC, Veersema S, Broekmans FJMM, Jansen FW, Schreuder Henk WRR (2014) Training of hysteroscopic skills in residency program : the Dutch experience. J Surg Educ 72(2):345–350. https://doi.org/10.1016/j.jsurg.2014.09.003

    Google Scholar 

  30. Elessawy M, Skrzipczyk M, Eckmann-scholz C et al (2016) Integration and validation of hysteroscopy simulation in the surgical training curriculum. J Surg Educ 74(1):84–90. https://doi.org/10.1016/j.jsurg.2016.06.007

    Google Scholar 

  31. Sleiman Z, Rassi E, Tanos V, Khazzaka A (2018) From the dry lab to the operating room : the first transferability study of hysteroscopic skills. Surg Endosc 33:1854–1857. https://doi.org/10.1007/s00464-018-6464-x

    Google Scholar 

  32. Munro MG, Advincula AP, Banks EH et al (2020) Essentials in minimally invasive gynecology manual skills construct validation trial. Obstet and gynecol 136(1):83–96. https://doi.org/10.1097/AOG.0000000000003936

    Google Scholar 

Download references

Acknowledgements

The authors thank Roxana Knobel, MD, PHD, for her support to build the simulator, and English teacher and translator Keila Campos de Oliveira for her linguistic support.

Statement of prior presentation or publications and/or abstract/poster presentation

The abstracts with partial results of this research were accepted for poster presentation at the 59 Brazilian Congress of Gynecology and Obstetrics in November 2019. These data have not been published elsewhere and the manuscript is not under review at any other journal.

Funding

None.

Author information

Authors and Affiliations

Authors

Contributions

ARPP: This author designed and managed the construction of the simulator, participated in the design of the study and in the construction of the measurement instruments, collected data, participated in statistical analyses, produced the manuscript, and approved its final version. LBG: This author participated in the design and construction of the measurement instruments, reviewed and rated the video recordings, and read and approved the final version of the manuscript. LKV: This author participated in the design and construction of the measurement instruments, reviewed and rated the video recordings, and read and approved the final version of the manuscript. GROF: This author participated in the design the study, in statistical analyses, and in the production of the manuscript and approved its final version.

Corresponding author

Correspondence to Ana Rita Peixoto Panazzolo.

Ethics declarations

Ethics approval and consent to participate

Approved by the IRB under CAAE - 88475118.7.0000.0121 on June 19, 2018, of the participating institutions. Written informed consent was obtained from participants before inclusion in the study.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Appendix

Appendix

Table 2 Technical performance checklist

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Panazzolo, A.R.P., Grando, L.B., Volpato, L.K. et al. The development and psychometric validation of a low-cost anthropomorphic 3D-printout simulator for training basic skills applicable to office-based hysteroscopy. Gynecol Surg 18, 7 (2021). https://doi.org/10.1186/s10397-021-01089-4

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s10397-021-01089-4

Keywords